ProCeas® the accurate analyzer for H2 purity control

Multicomponent and highly sensitive
Ease of use and fast response time
Low maintenance and no drift

You are here

Application note H2 purity control

Follow this link to view all

Our application notes

Why do you need to control impurities in hydrogen?

In a time when the world is looking for cleaner sources of energy, hydrogen is seen by many as an interesting alternative for cars, trains, etc.

The fast evolution of the hydrogen sector brings industrials to find new solutions to control the purity of hydrogen.

Quality of hydrogen is critical to insure that impurities does not affect the lifetime of fuel cells, engines, turbines. Gas such as sulfur and halogenated compounds can be problematic.

Technologies

ProCeas® hydrogen purity analyzer technology

The ProCeas® uses the patented OFCEAS technology, based on extractive absorption spectroscopy. This laser based technology is able measure gases down to sub-ppb levels of concentration, with an unforeseen quality of measurement in term of accuracy and repeatability.

The ProCeas® samples the gas at low pressure, with an internal pressure below 100 mbar absolute. This dilution reduces the volume needed to get a correct measurement, and optimizes the response time of the analyzer. In the end the system requires really small samples of hydrogen.

Finally, ProCeas® has an integrated auto calibration of the measurement system, allowing no drift under normal use of the instrument. ProCeas® is able to analyze a large range of impurities with limited service and maintenance costs.

Key features

Highly sensitive
Low detection level (sub-ppb)
Complete pre-calibrated laser infra-red spectrometer
No Zero drift and no Span drift – No daily calibration required
No nitrogen, zero air, purge or other carrier gas
Multigas analyser (H2S, CO, CO2, CH4, O2, NH3, H2O, HCHO, etc.)
Limited volume of sample needed for an analysis

Installation

Analytical specifications

What are the recommendations for maximum content of impurities?

ISO 14687-2 specifies the critical impurities and maximum levels to be detected in hydrogen.

ANALYTICAL SPECIFICATIONS
Component ISO 14687-2 limits (ppm)
 H2O 5
Total Hydrocarbons mainly CH4 2
O2 5
He 300
N2 / Ar 100
CO2 2
CO 0,2
Total sulfur compounds mainly H2S 0,004
HCHO 0,01
HCO2H 0,2
NH3 0,1
Total Halogenated compounds mainly HCL 0,05

ProCeas® Hydrogen purity

Is a complete pre-calibrated multicomponent (H2S, CO, CO2, CH4, H2O, O2, NH3, H2O, HCHO, etc.) laser infrared spectrometer for measurements of impurities in hydrogen.

ANALYTICAL SPECIFICATIONS
Component ISO 14687-2 Limits (ppm) LOD ProCeas®a(ppm)
 H2O 5 0,01
CH4 2 0,001
O2 5 1
CO2 2 0,2
CO 0,2 0,001
H2S 0,004 0,001
HCHO 0,01 0,001
HCO2H 0,2 0,005
NH3 0,1 0,001
HCL 0,05 0,001

a limit of detection 3 Sigma